Welcome back! In this week's tip for Empower Chromatography Data System (CDS), we continue our series on Peak Purity using both the PDA detector and QDa Mass detector and discuss reporting Peak Purity results. (In last week's tip, we discussed Multi Component Peak Purity, Tip #258).

Reporting of numerical and/or graphical results for Peak Purity is straight forward. We will focus on reporting only Peak Purity results. If you want to learn more about creating/working with Report Methods, refer to the Empower Tip series on Creating Report Methods (Tips #148 >#164)

Let's get started.

Step 1

If you are new to creating Report Methods, I recommend copying the default Report Methods from the PDA Defaults project into your working project. Highlight the Report Methods from PDA Defaults, right click and select 'Copy to Project'. Select your project, click 'OK' (figure 1).

figure 1

Step 2

One of those Report Methods, the PDA Purity Plot report, displays one Purity Plot for each peak. The Purity Angle and Purity Threshold will be displayed below each plot (figure 2).

figure 2

Step 3

Add Peak Purity results from the QDa Mass detector to our report by adding the Mass Analysis Plot to the report (figure 3).

figure 3

Step 4

In the properties of the Mass Analysis Plot, select 'Purity' on the Mass Analysis tab. You can limit the number of peaks plotted per page to make it easier to see the Purity View (figure 4).

figure 4

Step 5

Just as in Review, you can see the UV and MS spectra from the leading edge, apex, and trailing edge for each peak (figure 5).

figure 5

Step 6

If you do Multi Component Peak Purity as described in Tip #258, the '4 Pass Peak Purity Report' (also found in the PDA Defaults project) will display a purity plot for each Purity Pass for every peak (figure 6).

figure 6

Step 7

The plots described above can be found under PDA/MS Groups in the Report Publisher (figure 7).

figure 7

Step 8

Purity Angles and Purity Thresholds can be added to any existing report table by looking under PDA Peak (figure 8).

figure 8

It's that easy!

Final Note: This can be done with either the Pro or QuickStart interface.

Please rate this Empower Tip of the Week

Next time in Empower Tip #260- we will answer an Empower user question about working with QDa data.

JOIN US for an Empower: Ask Me Anything Webinar on January 20, 2022

Neil will answer your questions about Durable Data and Intrinsic Compliance.

Learn More and Register

JOIN US at [inform] 2022 | May 9-12, 2022 · Nashville, TN USA

Join the Waters Informatics community to discuss the latest forward-thinking digital innovations and technologies to elevate quality and improve laboratory effectiveness.

Learn More and Register

More Empower Resources:

Attachments

  • Original Link
  • Original Document
  • Permalink

Disclaimer

Waters Corporation published this content on 05 January 2022 and is solely responsible for the information contained therein. Distributed by Public, unedited and unaltered, on 05 January 2022 14:27:07 UTC.